We are here for you! +49 2602 9449-0 info@modul-technik.de

MS 1054

Application area: Intensive care
Product group: Media column

More free space through freestanding placement in the room

With our MS 1054 media column which can be installed freestanding in the room, we offer you a flexible system for all room situations in which wall-mounted supply systems are out of the question.

The core of the system is a compact media column which is firmly anchored to the floor and ceiling and in which all required media accesses such as medical gases, high voltage current, data and communications technologies are integrated.

The optionally available accessories make the MS 1054 a multi-purpose and flexible treatment station. Side-mounted stainless steel tubes incorporate infusion bottle holders, lighting elements and other aids. But storage consoles and drawer pull-outs can be installed at any height you want – ideal for every work situation and always exactly where they are required in the room.

Technical data

Equipment features
high-voltage technology
Read more…

Electric specifications
Nominal voltage: 230 V - 240 V / 50 Hz - continuous operation
Protection class: I
Protection type: IP 20

Equipment features
lighting technology
Read more…

Lighting technology LED downlight (optional)
Nominal voltage: 95 V – 260 V / 50 – 60 Hz
Protection class: II
Operating type: electronic converter (1-3,6 W)
Lamp output: 1 W
Protection type: IP 20

Equipment features
medical gas technology
Read more…

Operating pressure of medical gas technology
Oxygen: 5 bar
Compressed air: 5 bar
Vacuum: - 0,8 bar
Laughing gas 5 bar
CO2: 5 bar
AGSS: 5 bar

General information
Media current feed: From the top
Optional support tubes: Side-mounted, rigid or swivel
Total additional load: Max. 160 kg

Assembly, cleaning, maintenance and repair

Assembly

The medical supply units and single-bed lighting fixtures can be installed and mounted on solid or plasterboard construction walls. Customer provided substructure is required for plasterboard walls. Project-specific substructures are used for ceiling-mounted supply units.

Cleaning

Our products can be cleaned with conventional cleaning agents and disinfected with alcohol-free disinfectants.

Surfaces are protected by high-quality electrostatic powder coating. Non-coated parts are made of anodized aluminium or chrome-nickel steel. All plastic parts are resistant to cleaning agents and disinfectants.

Maintenance and repair

The system must be maintained for the first time after 5 years and then after every 2 years. A contractual service agreement can be concluded for the maintenance work if required. More information can be found in the respective operating instructions.

Accessories

Our comprehensive range of accessories means you can set up your work area exactly as you want it. Consult our Accessories Catalogue to find out about the wide range of options available to you.

Further products from product group Media column intensive care

No items were found.